3.2.3.2 Preset Setup
The Preset Setup window allows you to define temperature set point or sets of
temperature/target pairs to be used in the Absolute Mode Window and
Differential Mode Window.
To update the presets for Absolute
mode the current temperature control working
mode must be Absolute
, and for updating Differential presets, the working mode must
be Differential
(see Section 3.2.3.1: Mode).
To open the Preset Setup window (Differential Mode example):
1. From the Menu Window, press . The following window opens:
Figure 3-15: Preset Setup
To set a new list or update a predefined list:
2. Press . The PRESET button turns red.
3. Select a Preset No by using the
up/down arrows.
4. Set the temperature:
a. Press . The TEMP SET button turns red and the temperature
setting is enabled.
To cancel press again.
b. Press the TEMP SET numeric-edit field
. A floating Keypad opens.
c. Enter the required value by using the Keypad (for negative values, first type
the value and then the minus sign).
d. Press
to set the temperature and close the Keypad.
605-7040 Operation Manual
Rev D 3-16
June 2013