CI Systems SR-800R
7.2.2.2 Installation
1. Clean the Sensor port using the plastic straw (supplied with the new sensor):
a. Insert the plastic straw into the Sensor port.
b. Rotate the plastic straw several times in the Sensor port.
c. Remove the plastic straw from the Sensor port.
2. Apply a thin layer of thermal grease (supplied with the new sensor) on the bottom
section of the sensor (approx. 1 cm on the free end of the sensor), using a
toothpick (see Figure 7-6).
3. Insert the sensor into the port until the magnet secures it (see Figure 7-8).
4. Place the Sensor Locker and secure using the two screws (see Figure 7-8).
5. Connect the Sensor cable-connector (marked as EMT, see Figure 7-7).
6. Arrange the sensor cable.
605-7040 Operation Manual
Rev D 7-9
June 2013