CI Systems SR-800R
6.2.1 SETMODE
Summary: This is used to set the current temperature mode.
NOTE: It replaces the CM command.
Note: This command may take some time. Allow at least 1-2 seconds before
checking the mode with the GetMode Command.
S E T M O D E
Space
Mode
EOM
Input Operand: Mode - This is an integer value defining the temperature mode to
set the system:
1 = Absolute.
2 = Differential.
3 = Differential with Radiometric Offset Correction
Return Values: None
6.2.2 GETAXISSTATUS
Summary: This command retrieves the current motion status of the requested
Axis. This command may not indicate movement if called immediately after
calling the SetTarget, AxisHoming, etc. commands. Wait for about one (1) second
before calling this command.
G E T A X I S S T A T U S
Space
nAxisNumber
EOM
Input Operand: - NAxisNumber - A value from 1 to 4 representing the axis to be
accessed. If no Axis number is provided, the command will return the status of
combined axes in motion. This means that if any of the axes in the system are in
motion, then the result will be 1 (in Motion); otherwise 0, i.e., Stopped.
Return Values:
-1 = Error
0 = Stopped
1 = In Motion
2 = In Home Position (this is a momentary value when the axis is in motion and
is passing through Home).
605-7040 Operation Manual
Rev D 6-8
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents