CI Systems SR-800R
7.2.1.2 Installation
1. Apply a thin layer of thermal grease (supplied with the kit) on the bottom section
of the sensor (approx. 1cm on the free end of the sensor), using a toothpick:
Figure 7-6: Greasing the Sensor
2. Insert the sensor into the Emitter port until the magnet secures it.
3. Connect the Sensor male connector to the Blackbody (see Figure 7-3).
4. Arrange the sensor's wiring so that the installing the cover will not damage it.
5. Place the Sensor Cover and tighten the cover fasteners (see Figure 7-1).
NOTE
When placing the sensor cover ensure that there is no
obstruction and the cover closes easily. Do not use excessive
force.
605-7040 Operation Manual
Rev D 7-6
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com