- Press the DT SET numeric-edit field
.
The following floating Keypad opens:
Figure 3-10: Using the Floating Keypad
- Enter the desired temperature (by pressing the desired numbers) and press .
Results:
+ The Differential Mode Window re-opens.
+ The Te field continuously displays the temperature of the Blackbody Emitter.
+ The Tt field continuously displays the target temperature.
+ The DT SET button blinks until the Emitter temperature reaches the requested
temperature difference in the Te field.
2. Select a target for use from the default axis:
a. Press
(in some cases, the TARGET button can be labeled as
SOURCE or FILTER).
- The TARGET button turns red and the target selection is enabled.
b. Select a target by using the
up/down arrows. The target name is displayed
in the TARGET field.
c. Press
to execute your settings.
- The Target-Wheel starts to rotate toward the selected target.
- The TARGET button blinks until the selected target is valid.
d. To define the target parameters see Section 3.2.3.6.1: Motion Setup.
e. To select a target from another axis/Target-Wheel
see Section 3.2.3.6.2: Settings.
OR: Press the TARGET numeric-edit field
. The Motion Control
window opens (see Section 3.2.3.3: Motion Control). Select a Target and press
.
The Main Window re-opens.