CI Systems SR-800R
3.2.2 Differential Mode Window
(See Figure 3-9)
Upon pre-settings (see Section 3.2.3.1: Mode), the Differential Mode Window opens
immediately after powering-on the SR-800R and upon completion of the SR-800R
boot process (see Section 3.1: Powering-on the SR-800R):
Figure 3-9: Differential Mode Window
In the Differential Mode Window you can:
Set the temperature difference between the Blackbody Emitter and the target (DT
SET).
Select a target for use during tests (TARGET).
Select a pair of T SET and TARGET or only T SET from a predefined Preset
list (for detailed information on how to predefine the list, see Section 3.2.3.2:
Preset Setup).
To use the Differential Mode Window:
1. Set the temperature difference between the Blackbody Emitter and the target
(default selection):
a. Press
. The DT SET button turns red and temperature setting is
enabled.
b. To set a temperature use one of the following:
- Press the up/down arrows:
The
buttons increase/decrease the value by 0.1°.
The
buttons increase/decrease the value by 0.001°.
- Press
to execute your settings.
605-7040 Operation Manual
Rev D 3-9
June 2013