CI Systems SR-800R
3. To select a preset - (a combination of ∆T SET and TARGET, Source or Filter,
or only ∆T SET from a predefined Preset list (for detailed information on how to
predefine the settings, see Section 3.2.3.2: Preset Setup). The first four presets can
be selected directly from the Main Window, as follows:
a. From the Main Window (one of four): press one of the preset buttons
(P1 - P4; P1 in this example). The P1 button
turns red (a gray button -
, is invalid; no Preset is available):
Figure 3-11: Selecting a Preset Using the Preset Buttons
b. Press to execute your selection.
- The ∆T field continuously displays the temperature of the Blackbody
Emitter.
- The Tt field continuously displays the target temperature.
- The TARGET field is updated with the preset and the target is selected on
the Target Wheel.
- The P1 button blinks until the Emitter temperature reaches the requested
temperature in the DT SET field.
605-7040 Operation Manual
Rev D 3-11
June 2013