EasyManua.ls Logo

CI Systems ControlMaster SR-800R - Computer-To-GPIB Serial Poll Procedure; Description; Response

CI Systems ControlMaster SR-800R
185 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CI Systems SR-800R
6.2.43 Computer-to-GPIB Serial Poll Procedure
NOTE
This is a GPIB only command!
6.2.43.1 Description
The read status command indicates if long term operations have ended.
The response to this command is a single character. This character should be
interpreted bit-by-bit. The status byte can be read by querying the Controller serially.
This can be done at any time by the computer; although the Controller never initiates
a Service Request (SRQ).
6.2.43.2 Response
The Least Significant Bit (Bit #0) gives an indication of the Targets-Wheel (RAD)
status:
A value of "0" means that the wheel is rotating,
A value of "1" means that the wheel has reached its final position.
Bits #1, #2, and #4 are useable on the CTS-1 only.
Bit #3 gives temperature status: a value of "1" means that the set point temperature
has been reached.
All other bits are set to "0".
(LSB) 0 Target Wheel "0" = Rotating
1 RAD SYNC "1" = SYNC was found
2 CTS-1 Stage "0" = moving
3 Temperature Stable "1" = READY
4 CTS-1 Stage Home "1" = Home
5 --- Always "0"
6 Service Request Always "0"
(MSB) 7 --- Always "0"
605-7040 Operation Manual
Rev D 6-33
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com