f. Using the Keypad enter the starting temperature set point (dT Start > dT
Spec), and click .
- The floating Keypad closes.
- The entered value is displayed in the dT Start field:
Figure 3-23: MRTD Test Setup (4 of 5)
4. To add another set of target and temperature:
a. Press . Another line for the next target set is displayed under Step #1.
b. Repeat step 2 , replacing Step #1 with the next step number. A general view of
five steps/frequencies is displayed as shown below:
Figure 3-24: MRTD Test Setup (5 of 5)
5. To remove/delete the last line of the target press . The last line is deleted
(pressing Remove again will delete the last remaining line).
6. Press
to close and return to the MRTD main window.
605-7040 Operation Manual
Rev D 3-24
June 2013