CI Systems SR-800R
3.2.3.6.3 Communications
The Communications window allows you to set some defaults parameters of the SR-
800R Controller.
From the Options window, press
. The following window opens:
Figure 3-41: Communications Setup
1. To set the Control Mode:
a. This window allows you to select the preferred communication protocol
(RS232, Ethernet or GPIB) for remote control communication with a
computer.
If you choose the Ethernet (LAN) option, you must either obtain an IP
address automatically
from the Network server or define the IP address
"manually" by setting the IP address and SubNet Mask. The settings can be
changed by clicking on each three consecutive numbers/value and typing the
new number/value. You can also connect the SR-800R directly to a computer
(not through the local network). In this case you must use a "Direct LAN" or
crossover cable (this may be ordered from CI-Systems). You need also to
change the setting of your computer (the computer IP address should not
conflict with the SR-800R IP address - default is 192.168.0.2).
If you choose the GPIB (IEEE-488) option you also must choose the GPIB
address (1 to 15) by using the arrows (default address is 10).
b. To save the settings press
.
605-7040 Operation Manual
Rev D 3-39
June 2013