EasyManua.ls Logo

CI Systems ControlMaster SR-800R - Figure 3-26: MRTD Test Results Chart; Figure 3-27: MRTD Test Results Table

CI Systems ControlMaster SR-800R
185 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CI Systems SR-800R
5. Repeat steps 2 - 4 (up to 5 times). The MRTD is calculated using the following
equation:
()
(
)
(
)
2
NegativeTPositiveT
fMRTD
=
Upon completion, the following MRTD Test Results Chart is displayed:
Figure 3-26: MRTD Test Results Chart
6. You may press to view the MRTD Test results in a table format:
Figure 3-27: MRTD Test Results Table
7. You may stop running the MRTD test at any time by pressing .
To close and return to the last opened window (Figure 3-26) press
.
605-7040 Operation Manual
Rev D 3-26
June 2013
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com