b. Select the target with the smallest spatial frequency (T01 in this example),
and press the last available Step number (#1 in this example).
c. Press the dT Spec field
. A floating Keypad opens.
d. Using the Keypad enter a value related to the selected spatial frequency
according to the FLIR MRTD specifications, , and then press
.
- The floating Keypad closes.
- The value you entered is displayed in the dT Spec field.
e. Press the dT Start field
. A floating Keypad opens:
Figure 3-22: MRTD Test Setup (3 of 5)
605-7040 Operation Manual
Rev D 3-23
June 2013