6.2.37 RG
Summary: Use this command to return the absolute and differential temperature
ranges of the controller. The command returns the maximum/minimum
temperature pair of the current working mode of the controller (i.e., Absolute or
Differential).
Note: This is an old style SR-80 command that is used for backward compatibility
to CI SR80 BlackBody Controllers. Not recommended for use. See replacement
command, GetTemperatureRange.
Reference Command: GETTEMPERATURERANGE
RG
EOM
SR-80 command format compliance.
Input Operand: None
Return Values:
Return1 - Integer - Maximum Absolute/Differential Temperature
Return2 - Integer - Minimum Absolute/Differential Temperature