CI Systems SR-800R
Advantages of the removable BB are:
The removable probes are longer and are placed deeper into the emitter plate,
delivering even greater accuracy.
All other performance specifications for the new blackbodies will maintain their
high standards.
Here’s how simple it is (refer to Chapter 7 for step-by-step instructions on how to
replace a sensor):
1. Order a new calibrated sensor/s from CI Systems.
2. Replace the old sensor/s with the new one.
3. Upload the calibration data (via your PC).
Now your test station is back in operation - fully calibrated and NIST traceable.
However, users who prefer to carry out calibration themselves can do so using the
CK-800R Calibration Kit.
For more details about calibration methods, refer to Appendix A: Calibration
Methods.
605-7040 Operation Manual
Rev D 1-2
June 2013