CI Systems SR-800R
2. Offset:
To Insert Either Positive or Negative Global Positioning Offset:
a. Press the Offset field . A floating Keypad opens.
b. Using the Keypad enter the Offset value (in MSteps) and press . The
Keypad closes and the Offset value is displayed in the Offset field.
c. Press
to save your changes.
3. EFL (Effective Focal Length of the Collimator):
The controller automatically calculates the target angular size, or the spatial
frequency, of each 4-bar target in the system using the EFL value and the target
size (in mm). It is mainly used for MRTD tests.
To Insert the EFL Value of Your Optical Test Setup:
a. Press the EFL field
. A floating Keypad opens.
b. Using the Keypad enter your value (in mm) and press
. The Keypad
closes and the EFL value is displayed in the EFL field.
If the value is set to 0, all the target sizes will be displayed in units of mm;
otherwise, the target sizes are displayed in mRads or Cy/mRad.
c. Press
to save your changes.
4. Plate Distance:
The Plate Distance value indicates the distance between each of two consecutive
plates (measured in wheel MSteps). If each plate holds only one target in its
center, then the distance between targets is equal to the plate’s distance. The
controller automatically calculates and updates the Plates Distance parameter
according to the Plates value.
To update the Plates value:
a. Press the Plates field
. A floating Keypad opens.
b. Using the Keypad enter your value (in MSteps) and press . The
Keypad closes and the Plates value is displayed in the Plates field.
c. Press
to save your changes.
605-7040 Operation Manual
Rev D 3-29
June 2013