CI Systems SR-800R
4. To Change/Update the Size of an Existing Target:
a. Select a target from the list (you may browse the list by using the Up / Down
buttons).
b. Press the Size field
. A floating Keypad opens.
c. Using the Keypad enter the new target size (mm), and press
. The
floating Keypad closes, and the new target size is displayed in the Size field.
d. Press
, and confirm the modified Target Setup.
5. To Change/Update the Position of an Existing Target:
a. Select a target from the list (you may browse the list by using the Up / Down
buttons).
b. Press the Position field
. A floating Keypad opens.
c. Using the Keypad enter the new target position (MSteps), and press
.
The floating Keypad closes, and the new target position is displayed in the
Position field.
d. Press
, and confirm the modified Target Position.
e. To verify the target new positioning press and check that the target
wheel rotates to the desired location.
6. To Set the Temperature Offset: The temperature offset settings should be
carried out for every target. If the Temperature Offset = 0, then the system is
using the Radiometric Offset value, providing that the Radiometric Offset is set
to On (see Section 3.2.3.1, step 4 ):
a. In the Differential Mode window set the DT to 0 and select a target (see
Section 3.2.2).
b. Aim a FLIR towards the Focal Plane/Blackbody and check for differences. If
there are any, change the DT until no differences are observed.
c. Record the DT value.
d. Open the Target Setup window (see Figure 3-35).
e. Select the same target as selected in step a and enter the recorded DT value in
the Temperature Offset field (recorded in step c ).
f. Click
.
g. Repeat steps a - f for the entire targets.
605-7040 Operation Manual
Rev D 3-32
June 2013