Instructions for Use Fabius MRI SW 3.n 103
Operation
Starting Pressure Support Ventilation
The following examples and illustrations describe
starting Pressure Support ventilation from the
present ventilation mode “Volume Control”:
1 Press the »Pressure Support« key. The LED
associated with this key starts blinking. It
remains blinking until the selected mode of
operation is confirmed.
2 The Waveform window is replaced by the venti-
lator settings window and a message that pro-
vides instructions to confirm the mode change.
z If the ventilator settings are correct, confirm the
mode change.
z If the ventilator settings are not correct, for each
parameter that needs to change, press the cor-
responding soft key, select the correct value,
and confirm the change. When the parameter
changes are completed, confirm the ventilation
mode change.
z After the mode change is confirmed, the
»Pressure Support« key LED switches from
blinking to constantly on, the ventilator switches
to the Pressure Support mode, and the wave-
form is restored.