Instructions for Use Fabius MRI SW 3.n 103
Operation
Starting Pressure Support Ventilation
The following examples and illustrations describe
starting Pressure Support ventilation from the
present ventilation mode “Volume Control”:
1 Press the »Pressure Support« key. The LED
associated with this key starts blinking. It
remains blinking until the selected mode of
operation is confirmed.
2 The Waveform window is replaced by the venti-
lator settings window and a message that pro-
vides instructions to confirm the mode change.
z If the ventilator settings are correct, confirm the
mode change.
z If the ventilator settings are not correct, for each
parameter that needs to change, press the cor-
responding soft key, select the correct value,
and confirm the change. When the parameter
changes are completed, confirm the ventilation
mode change.
z After the mode change is confirmed, the
»Pressure Support« key LED switches from
blinking to constantly on, the ventilator switches
to the Pressure Support mode, and the wave-
form is restored.
564
2
1
521

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals