EasyManua.ls Logo

Dräger Fabius - Calibrate O 2 Sensor

Dräger Fabius
232 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for Use Fabius MRI SW 3.n 135
Configuration
Calibrate O
2 Sensor
z Press the »Calibrate O2 Sensor« key on the
Standby screen. The Standby soft key window
is replaced by instructions for performing the
calibration.
z Follow the instructions on the screen:
Remove O
2 sensor and expose to room air for
2 minutes.
To start O
2 Calibration press rotary knob.
z When the calibration begins, the instructions
are removed, and a »O
2 Calibration in
progress« message is displayed above the
Standby soft keys.
z When the calibration is completed, one of two
messages is posted above the Standby soft
keys: »O2 Sensor Calibration completed -
reinsert O
2 sensor« or »O2 Sensor Calibra-
tion Failed«.
Troubleshooting O
2 Sensor Calibration Failure
If the O
2 sensor cannot be calibrated, replace the
O2 capsule in the O2 sensor housing (see
page 189).
If the O
2 sensor still cannot be calibrated, call
DrägerService or your local authorized service
organization.
598
529
530
531
532

Table of Contents

Related product manuals