EasyManuals Logo

Dräger Fabius User Manual

Dräger Fabius
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #135 background imageLoading...
Page #135 background image
Instructions for Use Fabius MRI SW 3.n 135
Configuration
Calibrate O
2 Sensor
z Press the »Calibrate O2 Sensor« key on the
Standby screen. The Standby soft key window
is replaced by instructions for performing the
calibration.
z Follow the instructions on the screen:
Remove O
2 sensor and expose to room air for
2 minutes.
To start O
2 Calibration press rotary knob.
z When the calibration begins, the instructions
are removed, and a »O
2 Calibration in
progress« message is displayed above the
Standby soft keys.
z When the calibration is completed, one of two
messages is posted above the Standby soft
keys: »O2 Sensor Calibration completed -
reinsert O
2 sensor« or »O2 Sensor Calibra-
tion Failed«.
Troubleshooting O
2 Sensor Calibration Failure
If the O
2 sensor cannot be calibrated, replace the
O2 capsule in the O2 sensor housing (see
page 189).
If the O
2 sensor still cannot be calibrated, call
DrägerService or your local authorized service
organization.
598
529
530
531
532

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals