Instructions for Use Fabius MRI SW 3.n 39
Operation Concept
The Screen Display
The screen shows status, ventilation, and monitor-
ing data at a glance. The various screen pages use
largely the same layout to enable the user to find
information quickly.
1 Status bar displays the following status informa-
tion (from left to right):
Current ventilation mode
Time remaining in the alarm silence period
Status of reserve battery power
Current time
2 Alarm window which displays up to four of the
highest priority alarms.
3 Oxygen monitoring window which displays the
inspiratory oxygen concentration in percent (%)
along with high and low alarm limits
4 Respiratory volume monitoring window which
displays the patient’s respiratory rate in breaths
per minute (Freq), tidal volume (V
T), minute vol-
ume (MV), and the minute volume high and low
alarm limits
5 Breathing pressure monitoring window which
displays the patient’s positive end expiratory
pressure (PEEP), mean or plateau airway pres-
sure (MEAN or PLAT), and peak airway pres-
sure (PEAK) in cmH
2O (hPa) with high and low
alarm limits.
6 Breathing pressure trace window which dis-
plays a trace (waveform) of the patient’s breath-
ing pressure
7 Soft key labels
8 Flow Meter Monitor Window which displays a
graphical display of flow rates for O
2, AIR, and
N