Instructions for Use Fabius MRI SW 3.n 175
Cleaning
O
2 Sensor
Spirolog Infinity ID and SpiroLife flow sensors
Reprocess the flow sensor in accordance with the
corresponding Instructions for Use.
Sterilization
Strictly follow the correct Instructions for Use.
Steam sterilization at 273 °F (134 °C)
Strictly follow the correct Instructions for Use.
All components with the suitable reprocessing
options are listed in the maintenance list for
Fabius MRI components page 176. Follow the
hygiene regulations of the hospital!
CAUTION
Risk of equipment damage
The O
2 sensor must not be sterilized or disin-
fected!
CAUTION
The flow sensor must not be reprocessed in a
washer-disinfector. Do not clean with compressed
air, water jet, brush, etc. Otherwise, the thin wires
in the flow sensor may be destroyed. The Spirolog
and Infinity ID flow sensor must not be sterilized in
hot steam. The flow sensor is not resistant to high
temperatures and will be destroyed.
CAUTION
Only use clean disinfectant solutions to disinfect
the flow sensor. Contaminants, e.g., lint, may lead
to the destruction of the flow sensor.
WARNING
Risk of fire hazard
Allow the flow sensor to dry in air for at least
30 minutes after the use of disinfectants con-
taining flammable substances. These sub-
stances give off vapors that could ignite
during calibration.
CAUTION
The flow sensor can only be reused as long as
automatic calibration is possible.
CAUTION
Risk of damage to the medical device
The Spirolog and Infinity ID flow sensors must not
be sterilized.
CAUTION
Risk of damage to the medical device
SpiroLife flow sensors are not suitable for plasma
or radiation sterilization.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals