Getting Started
82
Instructions for Use Fabius MRI SW 3.n
Powering-Up the Machine
1 Turn the ON/OFF switch to the ON position.
When the ON/OFF switch is turned to the ON
position, the Fabius MRI performs extensive
self-tests on its internal hardware. As these
diagnostics are performed, each test and its
result appear on the screen. The result, Pass or
Fail, indicates the status of the tested compo-
nent.
We recommend the user to remain close to the
device within a range of up to four meters (13 feet),
to ascertain the verification by the acoustic tones of
the speakers.
During this self-test, two test tones are emitted to
show speakers functionality.
At the end of the self-diagnostics, one of three pos-
sible conclusions to the self-tests is posted on the
screen:
FUNCTIONAL
Every component of the monitoring system is in
satisfactory operational order. After a brief delay,
the Standby screen appears.
CONDITIONALLY FUNCTIONAL
A noncritical fault was detected. The Fabius MRI
may be used, but call DrägerService or your local
authorized service organization.
Press the rotary knob to continue operation.
CAUTION
The user must verify the acoustic tone are emitted
as the device can only verify the presence of the
speakers.
If no or only one tone is sounded, the device is
conditionally functional.
The complete loss of ventilation and monitoring
functionality might not be noticed.
Contact DrägerService.