Instructions for Use Fabius MRI SW 3.n 31
System Overview
Maximum Weight Loads
Component Maximum weight
Top of machine (with
adequate mounting
adapter only)
20 kg/44 lb
Drawers (each) 4 kg/8.8 lb
Writing table 25 kg/55 lb
Syringe/Infusion pump
(Note: mounted to
Dräger-supplied IV
pole)
9 kg/20 lb
WARNING
Do not place any object on this machine, writ-
ing table, or in the drawers unless it is specif-
ically labeled to be used in an MR scanning
room and on a Fabius MRI anesthesia system.
Objects placed on this machine that are not
designed for use with this anesthesia system
may be strongly attracted to the magnet and
may cause serious injury or death when the
machine is used in the MR scanning room.