Instructions for Use Fabius MRI SW 3.n 153
Configuration
Volume Alarms On/Off
1 Press the »Volume Alarms On« soft key in the
Setup window. The key label changes from
»Volume Alarms On« to »Volume Alarms
Off«, and the volume alarms are disabled.
Auto Set
2 Press the »Auto Set« soft key in the Setup win-
dow. The breathing pressure threshold is set to
4cmH
2O (hPa) below the current plateau pres-
sure data value.
NOTE
The »Volume Alarms On/Off« soft key label does
not appear in ManSpont mode because it is select-
able on the ManSpont screen.
NOTE
The threshold setting may not be less than
5cmH
2O (5 hPa) or greater than 30 cmH2O
(30 hPa).
NOTE
In the absence of a current plateau pressure data
value, pressing the soft key will have no effect.
NOTE
In SIMV/PS mode, the breathing pressure thresh-
old will be set relative to the mandatory ventilation
stroke.
503
1
2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals