Operation
112
Instructions for Use Fabius MRI SW 3.n
Ending Operation
1 Press and confirm the »Standby« key.
Monitoring and alarms are turned off and the venti-
lator stops.
2 Turn vaporizers off by turning the handwheel to
»0« until the button engages.
z Turn off fresh gas flow on the Fabius MRI.
Sleep mode will activate 2.5 minutes after fresh
gas is turned off.
z Close the cylinder valves.