Instructions for Use Fabius MRI SW 3.n 139
Configuration
Access Alarm Log
1 Press the »Access Alarm Log« key on the
Standby screen. The Standby soft key window
is replaced by the alarm log, which lists all
alarms with their dates and times.
z To scroll through the alarm log, turn the rotary
knob.
z To delete all alarms from the log, select and
confirm the »Clear Alarm Log« label.
z To exit the alarm log and return to the Standby
screen, select and confirm the return arrow.
Restore Site Defaults
2 Press the »Restore Site Defaults« key on the
Standby screen. The predefined site defaults
are restored, and a »Site Default settings
restored« message is displayed above the
Standby soft keys.
The site default settings are password protected.
Changes of the settings can be made via the
Standby Setup screen, see page 140.
CAUTION
Alarm log data is deleted if Fabius MRI is swit-
ched off or a total loss of electrical power supply
occurs.
WARNING
Risk of inappropriate ventilation settings
After site default settings have been restored
check whether the ventilation and monitoring
settings are appropriate to the patient.
561
534
561
535
1
2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals