Instructions for Use Fabius MRI SW 3.n 65
Assembly
Connecting the Breathing Hoses and
Breathing Bag
When connecting and disconnecting, always hold
the breathing hoses by the end sleeve, not by the
spiral reinforcement. Otherwise, the spiral rein-
forcement may be torn loose.
Breathing hoses with a damaged spiral reinforce-
ment can kink or become occluded.
Before each use, check the breathing hoses for
damage.
NOTE
Take care not to damage the breathing hoses.
WARNING
Risk of use of toxic or incompatible materials
Breathing hoses used on the Fabius MRI must
comply with current national standards.
WARNING
Do not use antistatic or conductive breathing
hoses or masks. There is a risk of burns in the
MRI environment or when using high fre-
quency electrosurgical equipment.
WARNING
Do not use silicone breathing hoses. They are
visible in MRI images and can be interpreted
as an artifact.
049

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals