Instructions for Use Fabius MRI SW 3.n 119
Monitoring
Sorting of displayed alarms
The alarms are sorted according to these catego-
ries. In the categories, the alarms are sorted and
displayed according to the internal priority system.
Priority 31 means the highest and priority 1 the low-
est priority. The priority numbers are shown in the
table "Fault-Cause-Remedy" on page 144.
A maximum of four alarms can be displayed in a list
at the same time. High-priority alarms are displayed
before low-priority alarms. Low-priority alarms are
sometimes only displayed after the cause for a
high-priority alarm has been remedied.
Example of audible alarm notification in the
event of several alarms.
If an alarm with the category Warning is currently
active and a new alarm with the category Warning
is generated, the alarm sequence starts again by
issuing an alarm with the same priority.
If alarm with the category Caution is also gener-
ated, no new audible alarm is issued. The Warning
alarm notification is not interrupted because it has
a higher priority.
If an alarm with the category Caution is currently
active and a new alarm with the category Caution is
generated, the alarm sequence starts again by
issuing an alarm with the same priority.
If an alarm with the category Caution is active and
an alarm with the category Warning is generated,
the alarm sequence is started for the warning
because the warning is assigned the highest prior-
ity.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min