Technical Data
212
Instructions for Use Fabius MRI SW 3.n
Classification II b
Conforming to Directive 93/42/EEC Appendix IX
UMDNS Code 10-134
Universal Medical Device Nomenclature System
Control Inputs Ranges
Low Oxygen Supply Pressure Alarm
APL-Valve MAN mode 5 to 70 cmH2O (hPa)
SPONT mode 1.5 cmH
2O (hPa)
accuracy from 5 to 15 L/min ±15 % of set value or ±3 cmH
2O (hPa) (greater
applies)
pressure drop at 30 L/min 3.4 cmH
2O (hPa) (wet and dry)
Alarm limit Warning signal when the pressure drops below 20 ±4 psi
(1.4 ±0.3 kPa x 100)
Alarm signal High priority alarm (Warning)
LED indicator The red LED indicator in the O
2 area of the gas flow control interface will
flash until the O2 supply is restored.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min