Technical Data
212
Instructions for Use Fabius MRI SW 3.n
Classification II b
Conforming to Directive 93/42/EEC Appendix IX
UMDNS Code 10-134
Universal Medical Device Nomenclature System
Control Inputs Ranges
Low Oxygen Supply Pressure Alarm
APL-Valve MAN mode 5 to 70 cmH2O (hPa)
SPONT mode 1.5 cmH
2O (hPa)
accuracy from 5 to 15 L/min ±15 % of set value or ±3 cmH
2O (hPa) (greater
applies)
pressure drop at 30 L/min 3.4 cmH
2O (hPa) (wet and dry)
Alarm limit Warning signal when the pressure drops below 20 ±4 psi
(1.4 ±0.3 kPa x 100)
Alarm signal High priority alarm (Warning)
LED indicator The red LED indicator in the O
2 area of the gas flow control interface will
flash until the O2 supply is restored.