Technical Data
204
Instructions for Use Fabius MRI SW 3.n
Electrical Safety Conformance
Conforms to:
z UL 60601-1
z IEC 60601-1
z CAN/CSA C22.2 No. 601.1-M90
General Safety Standards for
Anesthesia
z IEC 60601-2-13 plus US deviations
z ISO 8835-2
z EN 740
Freedom from latex
Fabius MRI is latex-free!
Latex-free breathing bags and breathing hoses
must be used for latex-free use.
Ventilator
Conforms to:
z ISO 8835-5
Control Inputs Ranges
P
MAX Pressure limiting 15 to 70 cmH2O (1 cmH2O resolution)
(15 to 70 hPa (1 hPa resolution))
(setting must be at least 10 cmH2O (10 hPa) above
PEEP; and in SIMV/PS mode, the PMAX setting must
also be greater than ΔP
PS+PEEP)
V
T Tidal volume 20 to 1400 mL (10 mL resolution)
V
T (SIMV/PS) Tidal volume 20 to 1100 mL (10 mL resolution)
f Breathing frequency 4 to 60 bpm (1/min) (1 bpm resolution)
(4 to 60 1/min (1/min resolution))
T
I/TE Inspiration/expiration ratio 4 : 1 to 1 : 4
T
IP/TI Inspiration pause 0 % to 50 % (1 % resolution)
PEEP End-expiratory pressure 0 to 20 cmH
2O (1 cmH2O resolution)
(0 to 20 hPa (1 hPa resolution))
P
INSP Inspiratory pressure 5 to 65 cmH2O (1 cmH2O resolution)
(5 to 65 hPa (1 hPa resolution))
(setting must be at least 5 cmH
2O (5 hPa) above
PEEP)