Technical Data
204
Instructions for Use Fabius MRI SW 3.n
Electrical Safety Conformance
Conforms to:
z UL 60601-1
z IEC 60601-1
z CAN/CSA C22.2 No. 601.1-M90
General Safety Standards for
Anesthesia
z IEC 60601-2-13 plus US deviations
z ISO 8835-2
z EN 740
Freedom from latex
Fabius MRI is latex-free!
Latex-free breathing bags and breathing hoses
must be used for latex-free use.
Ventilator
Conforms to:
z ISO 8835-5
Control Inputs Ranges
P
MAX Pressure limiting 15 to 70 cmH2O (1 cmH2O resolution)
(15 to 70 hPa (1 hPa resolution))
(setting must be at least 10 cmH2O (10 hPa) above
PEEP; and in SIMV/PS mode, the PMAX setting must
also be greater than ΔP
PS+PEEP)
V
T Tidal volume 20 to 1400 mL (10 mL resolution)
V
T (SIMV/PS) Tidal volume 20 to 1100 mL (10 mL resolution)
f Breathing frequency 4 to 60 bpm (1/min) (1 bpm resolution)
(4 to 60 1/min (1/min resolution))
T
I/TE Inspiration/expiration ratio 4 : 1 to 1 : 4
T
IP/TI Inspiration pause 0 % to 50 % (1 % resolution)
PEEP End-expiratory pressure 0 to 20 cmH
2O (1 cmH2O resolution)
(0 to 20 hPa (1 hPa resolution))
P
INSP Inspiratory pressure 5 to 65 cmH2O (1 cmH2O resolution)
(5 to 65 hPa (1 hPa resolution))
(setting must be at least 5 cmH
2O (5 hPa) above
PEEP)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals