Instructions for Use Fabius MRI SW 3.n 205
Technical Data
Insp Flow Inspiratory flow 10 to 75 L/min (1 L/min resolution) in Pressure Control
mode
10 to 85 L/min (1 L/min resolution) in PS and SIMV/PS
modes
ΔPPS (Pressure
Support)
Support Pressure 3 to 20 cmH2O (1 cmH2O resolution)
(3 to 20 hPa (1 hPa resolution))
ΔPPS (SIMV/PS) Support Pressure 3 to 20 cmH2O, OFF (1 cmH2O resolution)
(3 to 20 hPa, OFF (1 hPa resolution))
Freq Min Apnea Ventilation minimum
frequency
3 to 20 bpm (1 bpm resolution) and “OFF”
(3 to 20 1/min (1/min resolution) and “OFF”)
Trigger Trigger Level 2 to 15 L/min (1 L/min resolution)
T
INSP SIMV Inspiratory Time 0.3 to 4.0 s
Pressure Support Ventilation Mode
The Pressure Support Ventilation mode has been verified under the following range of simulated patient
conditions:
Endotracheal Tube Size: 0.18 in to 0.32 in (4.5 mm to 8 mm)
Patient Lung Compliance: 10 mL/cmH
2O to 100 mL/cmH2O
(10 mL/hPa to 100 mL/hPa)
Unassisted Patient Tidal Volume: 50 mL to 1000 mL
Patient Breath Rate (bpm) (1/min): 10 to 35
Delivery Accuracy
P
MAX Pressure limiting ±5 cmH2O (±5 hPa) of setting
V
T Tidal volume ±5 % of setting or 20 mL, whichever is greater
discharged to atmosphere, no compliance compensa-
tion)
f Breathing frequency ±1 bpm (1/min) of setting
T
I/TE Inspiration/expiration ratio ±5 % of setting
T
IP/TI Inspiration pause ±25 % of setting
PEEP End-expiratory pressure ±2 cmH
2O (±2 hPa) or ±20 % of setting, whichever is
greater