Assembly
70
Instructions for Use Fabius MRI SW 3.n
Connecting the Flow Sensor
1 Connect the volume line cable onto the connec-
tor on the underside of the breathing system.
2 Connect the volume line cable to the port
labeled on the connector panel on the back
of the machine.
Connecting the APL Bypass and
PEEP/P
MAX Hoses
3 Plug the control hose to the connection port on
the PEEP/P
MAX valve and to the connection
port marked PEEP on the connection panel.
4 Plug the control hose to the connection port on
the APL Bypass valve and to the connection
port marked APL on the connection panel.
NOTE
The APL bypass hose is larger than the
PEEP/PMAX hose.
019
1
2
020
3
4