Instructions for Use Fabius MRI SW 3.n 203
Technical Data
Recommended separation distances
Information regarding separation distances
(IEC 60601-1-2: 2001, tables 205 and 206)
Recommended separation distances between portable and mobile RF telecommunication
devices and the Fabius MRI
max.
P
EIRP (W)
3 V/m
distance
1)
(m)
1) 3 V/m distance to transmitters with frequencies from 150 kHz to 2.5 GHz, otherwise 1 V/m distance.
1 V/m
distance
1)
(m)
Note
0.001 0.06 0.17
0.003 0.10 0.30
0.010 0.18 0.55
0.030 0.32 0.95 e.g. WLAN 5250 / 5775 (Europe)
0.100 0.58 1.73 e.g. WLAN 2440 (Europe), Bluetooth
0.200 0.82 2.46 e.g. WLAN 5250 (not in Europe)
0.250 0.91 2.75 e.g. DECT devices
1.000 1.83 5.48 e.g. GSM 1800- / GSM 1900- / UMTS- mobiles,
WLAN 5600 (not in Europe)
2.000 2.60 7.78 e.g. GSM 900 mobiles
3.000 3.16 9.49