Technical Data
202
Instructions for Use Fabius MRI SW 3.n
radiated RF
(IEC 61000-4-3)
80 MHz to 2.5 GHz:
10 V/m
10 V/m Recommended separation distance
from portable and mobile RF transmit-
ters with transmission power P
EIRP to
the Fabius MRI including its lines:
1.84 m *
√PEIRP
1)
RF coupled into
lines
(IEC 61000-4-6)
150 kHz to 80 MHz:
10 V within ISM bands,
10 V Recommended separation distance
from portable and mobile RF transmit-
ters with transmission power P
EIRP to
the Fabius MRI including its lines:
1.84 m *
√PEIRP
1)
3 V outside ISM bands
2)
3 V
1) For PEIRP the highest possible “equivalent isotropic radiated power” of the adjacent RF transmitter has to be inserted
(value in Watt). Also in the vicinity of equipment marked with the symbol interference may occur. Field strengths from
fixed, portable or mobile RF transmitters at the location of the Fabius MRI should be less than 3 V/m in the frequency
range from 150 kHz to 2.5 GHz and less than 1 V/m above 2.5 GHz.
2) ISM bands in this frequency range are: 6.765 MHz to 6.795 MHz, 13.553 MHz to 13.567 MHz, 26.957 MHz to 27.283 MHz,
40.66 MHz to 40.70 MHz.
Information regarding electromagnetic immunity
(IEC 60601-1-2: 2001, tables 202, 203, 204)
Immunity against IEC 60601-1-2 test level Compliance
level (of this
equipment)
Electromagnetic environment