Instructions for Use Fabius MRI SW 3.n 113
Operation
End of Vaporizer Use
If the Vapor is not going to be used for up to six
months, it may remain filled;
If the Vapor is not going to be used for more than
6 months, see "Shut-Down" in separate Instruc-
tions for Use provided with the vaporizer.
If the Vapor remains on the machine:
z For intervals of more than one week, anesthetic
agent loss from the vaporizer chamber can be
minimized by using the »T« setting.
z The locking lever on the plug-in adapter should
remain in the left (locked) position.
z Keep Vapor within the permissible temperature
range (see separate Instructions for Use pro-
vided with the vaporizer
z Observe expiration date of anesthetic agent.
If the Vapor does not remain on the machine:
z see "Disconnecting the Vapor" and "Transport
of Filled Vapors" in separate Instructions for
Use provided with the vaporizer.
When Fabius MRI is not in use
If Fabius MRI is not used for an extended period:
z Unplug the medical gas hoses from the wall
supply points of the central gas supply.
z Close the cylinder valves on the reserve gas
cylinders.
WARNING
Damage of materials and health risks
Never switch off fresh gas flow before the
vaporizer is switched off. A vaporizer must
never be left switched on without a fresh gas
flow, because high-concentration anesthetic
vapor may leak into machine lines and ambi-
ent air.
NOTE
Leave the Fabius MRI plugged into mains power in
order to charge the battery.