Assembly
50
Instructions for Use Fabius MRI SW 3.n
Activating the Battery
The Fabius MRI anesthesia machine is shipped
with the battery fuse disconnected in order to pre-
vent discharge during shipment and storage prior to
installation.
z Remove the battery fuse from the top drawer of
the machine.
z Remove the battery fuse from its packaging.
1 Insert the battery fuse into the battery fuse
holder. Turn the fuse holder 1/4-turn clockwise
until it is snug.
WARNING
For operation in magnetic fields the combina-
tion of vaporizers, anesthesia workstation,
and MRI scanner must be tested by experts
(trained and factory authorized technical ser-
vice representatives for anesthetic machines
and MRI scanners) prior to the first use to
ensure proper vaporizer and interface func-
tion in the specific magnetic fields. Otherwise
uncontrolled concentrations, leakage, and/or
malfunction of the interlock system may
occur. The testing has to take into consider-
ation all positions of the anesthesia worksta-
tion, including the vaporizer, in which it will
operate in the MRI environment during daily
use. Additionally, it is necessary to check if
the imaging of the MRI scanner is adversely
affected by the vaporizer and the anesthesia
workstation.
010
1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals