Instructions for Use Fabius MRI SW 3.n 75
Assembly
Auxiliary power outlets
In cases of a blown safety fuse for the auxiliary out-
lets:
Remedy the fault
Have the safety fuse replaced by an electrician
WARNING
The auxiliary outlets are not powered by the
uninterruptible power supply (UPS) in the
event of a power failure.
WARNING
Risk of electric shock
Connecting equipment to the auxiliary power
outlets may cause the patient leakage current
to rise above the permitted values if a protec-
tive earth conductor should fail. The risk of
electric shock cannot be excluded in such
cases.
WARNING
Do not connect life-supporting devices to the
auxiliary power outlets of the anesthesia
workstation. The auxiliary power outlets are
not powered by the uninterruptible power sup-
ply (UPS) in the event of a power failure.
WARNING
Risk of increased leakage current
Do not connect HF surgical devices to the
auxiliary power outlets.
WARNING
Risk of increased leakage current
Additional power adapter outlets must not be
connected to the auxiliary power outlets.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals