Instructions for Use Fabius MRI SW 3.n 127
Monitoring
Volume Monitoring Alarms
While the ventilator is on and the volume alarms are
enabled, apnea alarms are generated if the respira-
tory volume monitor does not sense a valid breath
for a specified period (see »APNEA FLOW« on
page 162).
While the ventilator is off and the system is in Man-
Spont mode, these alarms are generated at
30 seconds (Caution) and 60 seconds (Warning).
The Fabius MRI's volume alarms are automatically
enabled when the ventilator is switched from
Standby to a ventilation mode.
Setting Minute Volume Alarm Limits
Follow the procedure "Setting Alarm Limits" on
page 121 to change the high or low alarm limit.
Disabling Volume Alarms
Visual and audible volume alarms can be turned on
or off during operation using the »Setup« key. See
“Volume Alarms On/Off” on page 153.