174 Instructions for Use Fabius MRI SW 3.n
Cleaning
Removing anesthetic gas scavenging
system (AGSS)
z Remove the anesthetic gas receiving system
(AGS) including scavenging hose and exhaust
hose from the medical device.
Removing the Suction System
z Remove suction bottle assembly, including bot-
tle and regulator.
Reprocessing Breathing system
All parts of the respiratory system, ventilator
diaphragm, Y-piece, respiratory hoses, respira-
tory bag, parts of the absorber, parts of the sec-
retion suction unit and parts of the anesthetic
gas scavenging system.
z Thermally disinfect in the cleaning/disinfection
machine at 200 °F (93 °C) /10 minutes.
Only use neutral or mild alcalic cleaning agents
(e.g. Neodisher Medizym, Neodisher FA) and
completely demineralized water!
When using thermal disinfection, the addition of
disinfectant chemicals is not necessary; risk of
corrosion!
WARNING
Risk of infection
Always wear protective gloves when emptying
the suction bottle.
Observe the hospital hygiene regulations.
NOTE
Reprocessing/disinfection instructions for the
reusable suction bottle and the suction regulator,
refer to their respective Instructions for Use.
WARNING
After washing, a hot steam sterilization is
required to completely dry the breathing sys-
tem.
Insufficient drying of the control areas found
in the valve plate can lead to impairment of
device function or to failure of the medical
device.