Technical Data
200
Instructions for Use Fabius MRI SW 3.n
Electromagnetic emissions
The Fabius MRI is intended for use in the electro-
magnetic environment specified below. The user of
the Fabius MRI should assure that is used in such
an environment.
Information regarding electromagnetic emissions
(IEC 60601-1-2: 2001, table 201)
Emissions Compliance
according to
Electromagnetic environment
RF emissions (CISPR 11) Group 1 The Fabius MRI uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
Class B The Fabius MRI is suitable for use in all
establishments including domestic establish-
ments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic pur-
poses.
Harmonic emissions
(IEC 61000-3-2)
Class A
Voltage fluctuations / flicker
(IEC 61000-3-3)
Complies