Cleaning
172
Instructions for Use Fabius MRI SW 3.n
Removing the Waste Gas Port
z Unscrew the waste gas port.
Removing the Flow Sensor
1Loosen fitting on the expiration port.
2 Remove the flow-sensor guard.
3 Extract the flow sensor.
CAUTION
Risk of flow measurement failure
Disinfecting or cleaning the flow sensors by
machine will damage them and cause the flow
measurement to fail.
Disinfect and clean the flow sensors as described
in the Instructions for Use of the Spirolog, Infinity
ID flow sensor and SpiroLife flow sensors.
CAUTION
Risk of flow measurement failure
Sterilizing the Spirolog and Infinity ID flow sensors
in high-temperature steam will damage them and
cause the flow measurement to fail.
Disinfect and clean the flow sensor as described
in the Instructions for Use of the Spirolog, Infinity
ID flow sensor and SpiroLife flow sensors.