Cleaning
172
Instructions for Use Fabius MRI SW 3.n
Removing the Waste Gas Port
z Unscrew the waste gas port.
Removing the Flow Sensor
1Loosen fitting on the expiration port.
2 Remove the flow-sensor guard.
3 Extract the flow sensor.
CAUTION
Risk of flow measurement failure
Disinfecting or cleaning the flow sensors by
machine will damage them and cause the flow
measurement to fail.
Disinfect and clean the flow sensors as described
in the Instructions for Use of the Spirolog, Infinity
ID flow sensor and SpiroLife flow sensors.
CAUTION
Risk of flow measurement failure
Sterilizing the Spirolog and Infinity ID flow sensors
in high-temperature steam will damage them and
cause the flow measurement to fail.
Disinfect and clean the flow sensor as described
in the Instructions for Use of the Spirolog, Infinity
ID flow sensor and SpiroLife flow sensors.
092
063
12
3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals