Instructions for Use Fabius MRI SW 3.n 207
Technical Data
Anesthesia Gas Supply Module
Fresh Gas Flow Indicators: .
O
2, N2O, AIR: Range and accuracy: 0.0 to 12.0 L/min ±10 % of reading or
0.12 L/min into an ambient atmosphere of 14.7 psi
(1.013 kPa x 100) at 20 °C (68 °F).
Resolution: 0.1 L/min
Fresh Gas Flow Stability:
O
2 and N2O: ±10 % of setting with pipeline pressures between 45 to 65 psi (3.1 to 4.5 kPa x 100)
AIR: ±10 % of setting with pipeline pressures between 50 to 55 psi (3.4 to 3.8 kPa x 100)
Air flow rate will vary proportionally with supply pressures outside 50 to 55 psi (3.4 to 3.8 kPa x 100).
Total Fresh Gas Flowmeter:
Range and accuracy: 0 to 10 L/min ±10 % of full scale at STP,
calibrated with 50 % O
2/ 50 % N2O gas mixture
0 to 10 L/min ±15 % of full scale at STP for all other gas mix-
tures
Resolution: 0.5 L/min from 0.5 to 2 L/min
1.0 L/min from 2 to 10 L/min
O
2 flush (bypass): at 55 psi (3.8 kPa x 100): max. 50 L/min
at 50 psi (3.4 kPa x 100): min. 35 L/min
Common Gas Outlet Pressure Limit: 13 psi (0.9 kPa x 100), maximum
Auxiliary Oxygen Flowmeter
Connection Staged connector for use with different hose diameters
Fresh-gas flow 0 to 10 L/min
Accuracy ±5 % of full scale
Resolution 0.5 L/min