Instructions for Use Fabius MRI SW 3.n 207
Technical Data
Anesthesia Gas Supply Module
Fresh Gas Flow Indicators: .
O
2, N2O, AIR: Range and accuracy: 0.0 to 12.0 L/min ±10 % of reading or
0.12 L/min into an ambient atmosphere of 14.7 psi
(1.013 kPa x 100) at 20 °C (68 °F).
Resolution: 0.1 L/min
Fresh Gas Flow Stability:
O
2 and N2O: ±10 % of setting with pipeline pressures between 45 to 65 psi (3.1 to 4.5 kPa x 100)
AIR: ±10 % of setting with pipeline pressures between 50 to 55 psi (3.4 to 3.8 kPa x 100)
Air flow rate will vary proportionally with supply pressures outside 50 to 55 psi (3.4 to 3.8 kPa x 100).
Total Fresh Gas Flowmeter:
Range and accuracy: 0 to 10 L/min ±10 % of full scale at STP,
calibrated with 50 % O
2/ 50 % N2O gas mixture
0 to 10 L/min ±15 % of full scale at STP for all other gas mix-
tures
Resolution: 0.5 L/min from 0.5 to 2 L/min
1.0 L/min from 2 to 10 L/min
O
2 flush (bypass): at 55 psi (3.8 kPa x 100): max. 50 L/min
at 50 psi (3.4 kPa x 100): min. 35 L/min
Common Gas Outlet Pressure Limit: 13 psi (0.9 kPa x 100), maximum
Auxiliary Oxygen Flowmeter
Connection Staged connector for use with different hose diameters
Fresh-gas flow 0 to 10 L/min
Accuracy ±5 % of full scale
Resolution 0.5 L/min

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals