Technical Data
206
Instructions for Use Fabius MRI SW 3.n
High Pressure Safety Relief Valve
75 ±5 cmH
2O (75 ±5 hPa)
Negative Pressure Safety Relief Valve (Ambient
AIR Inlet Valve)
–7.5 to –9 cmH2O (–7.5 to –9 hPa)
Minimal Limit Pressure
–8.5 cmH2O (–8.5 hPa)
System Compliance Compensation Measurement
0.2 to 6.0 mL/cmH2O (0.2 to 6.0 mL/hPa) ±0.2 mL/cmH2O (±0.2 mL/hPa or ±10% of actual
compliance, whichever is greater