Technical Data
210
Instructions for Use Fabius MRI SW 3.n
Breathing System
Volume with reusable absorber
incl. filled absorber, excl. hoses 1.7 L + bag
Volume with Drägersorb CLIC adapter
incl. filled absorber, excl. hoses 1.7 L + bag
excl. absorber, excl. hoses 0.8 L + bag
Volume absorber
reusable absorber, filled 1.5 L
Clic absorber
(Drägersorb CLIC Free)
1.2 L
Compliance
(in automatic ventilation modes, e.g. Volume Control, excluding patient hoses)
with reusable absorber 0.35 mL/cmH
2O (0.35 mL/hPa)
with Drägersorb CLIC adapter,
including absorber
0.35 mL/cmH
2O (0.35 mL/hPa)
with Drägersorb CLIC adapter,
excluding absorber
0.35 mL/cmH2O (0.35 mL/hPa)
Resistance
(Reusable absorber or CLIC absorber with or without flexible bag arm, normal operation, filled with
Drägersorb 800 +)
Inspiratory Expiratory
As per ISO 8835-2, dry, max.
±6 hPa (±6 cmH
2O), with hose
set for adults M30146
–4.4 cmH
2O
(–4.7 hPa)
4.2 cmH2O
(4.4 hPa)
As per ISO 8835-2, dry, sole bre-
athing system without
patient hoses
–3.7 cmH
2O
(–3.7 hPa)
3.7 cmH2O
(3.7 hPa)
Typical leak <50 mL/min