Instructions for Use Fabius MRI SW 3.n 209
Technical Data
O
2 Cell Measurement Performance
Response time Less than
16 seconds
measured values are not pressure
compensated
Warm-up time after 5 min error 3 % of measured value
Drift Sensitivity ±1 % of measured value / 8h
Cross Sensitivity 1 vol.% O
2 at 70 vol.% N2O and 5 vol.% CO2
with 4 vol.% Halothane
or with 5 vol.% Enflurane
or with 5 vol.% Isoflurane
or with 10 vol.% Sevoflurane
effect of humidity / sensitivity max. ±0.02 % of measured value / % relative humidity
Service life of O
2 sensor cell >12 months at 25 °C (77 °F), 50 % relative humidity, 50 % O2
gas mixture (or >5000 hours at 100 Vol% O2)