Operation
116
Instructions for Use Fabius MRI SW 3.n
1 Remove the pipeline supply hoses from the
central supply.
z Press the O
2 flush to depressurize the entire
system.
Moving the Machine
The Fabius MRI has a central caster brake bracket
on the front of the machine. To unlock the castors,
release the brake by lifting the bracket (e.g., with
the tip of your foot).
CAUTION
Do not attempt to move the anesthesia machine
while the casters are locked.
CAUTION
Watch hoses connected to the patient while mov-
ing the Fabius MRI.
CAUTION
Do not push or pull the Fabius MRI using the
absorber system, vaporizers, ventilator, or moni-
tor.
CAUTION
The Fabius MRI should be moved by people who
can handle its weight (approximately 425 lb
(192.8 kg)). Take extra care to avoid tipping when
moving on ramps, around corners, and over
raised thresholds such as door frames and eleva-
tors.
013
1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals