Instructions for Use Fabius MRI SW 3.n 93
Operation
CLIC Adapter (Optional)
The disposable CLIC adapter absorber can also be
used on the Fabius MRI. For information on install-
ing the CLIC adapter, consult its Instructions for
Use.
Points to note when using the CLIC adapter
with neonates and infants
The compliance of the breathing system is reduced
when the disposable absorber is removed. As a
result, the ventilator delivers a higher tidal volume
to the patient during IPPV/SIMV and the airway
pressure increases accordingly. The effect on the
tidal volume and airway pressure is negligible in
adults with a lung compliance of approx.
50 mL/cmH
2O. In neonates and infants with a lung
compliance of approx. 5 mL/cmH
2O or less, the
tidal volume and airway pressure can increase con-
siderably and harm the patient if the disposable
absorber is removed during IPPV/SIMV.
Before removing the disposable absorber:
z Set P
MAX to the plateau pressure.
As soon as the absorber has been
replaced:
z Set P
MAX to the original value again.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals