Assembly
58
Instructions for Use Fabius MRI SW 3.n
Connecting the Reserve Gas Cylinders
for N
2O, AIR and O2 ( for Pin-Index
Mounting)
WARNING
Use only non-magnetic (e.g., aluminum) cylin-
ders with the Fabius MRI. Steel cylinders will
cause serious injury or death if brought into
an MR scanning room.
WARNING
When attaching a cylinder, ensure that only
one washer is installed between the cylinder
and the yoke gas inlet. The use of multiple
washers will inhibit the pin-index safety sys-
tem. Be sure to verify the presence of the
index pins each time a cylinder is installed.
Never attempt to override the pin-index safety
system.
WARNING
Do not oil or grease the O2 cylinder valves and
O
2 pressure regulator. There is a risk of explo-
sion.
CAUTION
Even if the gas supply is connected to a medical
gas pipeline, the cylinders should remain on the
device in reserve.
NOTE
If cylinder valves are leaky or difficult to open or
close, they must be repaired in accordance with
the manufacturer’s specifications.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals