Instructions for Use Fabius MRI SW 3.n 187
Maintenance
Safety Checks
The safety checks are no substitute for the preven-
tive maintenance measures (including preventive
replacement of wear parts) indicated by the manuf-
acturer.
1 Check accompanying documents:
– latest Instructions for Use are available
2 Verify that the device combination is in good
condition:
– all labels are complete and legible
– there is no visible damage
– Fuses which are accessible from the outside
are in compliance with the specified values
3 Check that the equipment of the medical device
is complete according to the Instructions for
Use.
4 Check the electrical safety according to
IEC62353
5 Check safety features:
– Check correct functioning of the alarm genera-
tor.
– Check correct functioning of the alarm genera-
tion of lack of O
2.
– Check correct functioning of O
2 measurement.
– Check correct functioning of flow measurement.
– Check correct functioning of P
AW, PEEP, APL
and P
MAX pressure measurement.
– Check correct functioning of power fail alarm
and battery backup function.
– Check correct functioning of Vaporizer Interlock
function.
– Check correct functioning of auxillary air supply
and safety valves of ventilator.
– Check correct functioning of S-ORC.
– Check correct functioning of vaporizers accor-
ding the IfU of vaporizers.
CAUTION
The safety checks must be carried out in the spe-
cified intervals. Otherwise, the correct functioning
of the medical device can be impaired.