Instructions for Use Fabius MRI SW 3.n 187
Maintenance
Safety Checks
The safety checks are no substitute for the preven-
tive maintenance measures (including preventive
replacement of wear parts) indicated by the manuf-
acturer.
1 Check accompanying documents:
latest Instructions for Use are available
2 Verify that the device combination is in good
condition:
all labels are complete and legible
there is no visible damage
Fuses which are accessible from the outside
are in compliance with the specified values
3 Check that the equipment of the medical device
is complete according to the Instructions for
Use.
4 Check the electrical safety according to
IEC62353
5 Check safety features:
Check correct functioning of the alarm genera-
tor.
Check correct functioning of the alarm genera-
tion of lack of O
2.
Check correct functioning of O
2 measurement.
Check correct functioning of flow measurement.
Check correct functioning of P
AW, PEEP, APL
and P
MAX pressure measurement.
Check correct functioning of power fail alarm
and battery backup function.
Check correct functioning of Vaporizer Interlock
function.
Check correct functioning of auxillary air supply
and safety valves of ventilator.
Check correct functioning of S-ORC.
Check correct functioning of vaporizers accor-
ding the IfU of vaporizers.
CAUTION
The safety checks must be carried out in the spe-
cified intervals. Otherwise, the correct functioning
of the medical device can be impaired.