Operation
90
Instructions for Use Fabius MRI SW 3.n
Low Flow Anesthesia
With low flow anesthesia (means flow 1.0 L/min)
the condensation of patients humidity from the
expired breath is natural. Condensation occurs in
the hoses. To avoid water accumulating in the
hoses, a water trap should be integrated in the ven-
tilator hose. In long term low flow anesthesia the
additional use of watertraps in the expiratory hose
is recommended. Empty water traps if their water
level exceeds the maximum water level limit.
Nitrogen Wash-Out (If Required)
During anesthesia induction, air containing about
77 % nitrogen (N
2) remains in the compact breath-
ing system (and in the patient's lungs). If the unit will
be used for a low-flow anesthesia case, press the
O
2 Flush to remove this N2.
Replacing CO2 Absorbent
The CO2 absorbent in the compact breathing sys-
tem should be replaced when two-thirds of the CO2
absorbent has changed color. Dräger recommends
the use of Drägersorb 800 Plus or Drägersorb
FREE. The color change indicates that the CO
2
absorbent can no longer absorb CO
2
(Drägersorb 800 Plus and Drägersorb FREE
change from white to violet).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals