Instructions for Use Fabius MRI SW 3.n 91
Operation
WARNING
Do not flush dry gas continuously for unne-
cessarily long periods through the soda lime
in the anesthesia system!
Otherwise the soda lime will dehydrate. If the
moisture level falls below a minimum level,
undesirable reactions generally occur, regard-
less of the type of soda lime and the inhalation
anesthetic used:
–reduced CO2 absorption,
increased heat generation in the absorber
and therefore increased breathing gas
temperature,
CO formation,
absorption and/or breakdown of the inha-
lationanesthetic.
The above mentioned reactions may endanger
the patient, e.g.:
CO poisoning
insufficient depth of anesthesia
burns of the airway.
CAUTION
Soda lime irritates the skin and there is a risk of
serious damage to eyes.
If soda lime has leaked:
Powdered soda lime must not be inhaled or
swallowed.
Put on protective gloves and goggles, or a
face mask.
In case of coming in contact with eyes, rinse
immediately with large amounts of water and
consult a physician immediately, otherwise it
may lead to eye damage.
Powdered soda lime on the skin must be
washed off immediately, because it may irri-
tate the skin.
NOTE
Please refer to the specific Instructions for Use for
"Drägersorb 800 Plus or Drägersorb FREE".

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius and is the answer not in the manual?

Dräger Fabius Specifications

General IconGeneral
TypeAnesthesia Workstation
DisplayColor TFT
Gas SupplyO2, N2O, Air
CO2 AbsorberYes
ManufacturerDräger
ModelFabius
Ventilation ModesSIMV
Inspiratory Time0.2-5 s
PEEP0 to 20 cmH2O
Vaporizer Mounts2
O2 Flow0 to 15 L/min
Air Flow0 to 15 L/min

Related product manuals